Title:
|
A Comparative Study on Accuracy of Anterior Chamber Depth Measured with OrbScan and Pentacam
|
Author(s):
|
Ghandehari-Motlagh Mohammad, MirJalili Nafise
|
Presentation Type:
|
Poster
|
Subject:
|
Cornea and Anterior Segment
|
Others:
|
|
Presenting Author:
|
|
Name:
|
Mohammad Ghandehari Motlagh
|
Affiliation :(optional)
|
Sadra Eye Surgery Center ,Isfahan-Iran
|
E mail:
|
oculist233@aol.com
|
Phone:
|
031-36689608
|
Mobile:
|
09132000891
|
|
|
Purpose:
|
Comparing anterior chamber depth measured with OrbScan and Pentacam compared with the Smith method results.
|
Methods:
|
Smith method (since 1979) is a reliable method of measuring ACD only with help of slit lamp. In this study 102 phakic eyes as PRK candidates were imaged with both OrbScan and Pentacam and finally ACD was measured thru Smith method with slit lamp. ACD measured with Smith method was presumed as the gold standard and was compared with ACD of the 2 imaging devices. Contraindication cases for PRK and pseudophakic eyes have been excluded from the study.
|
Results:
|
Mean age of the patients was 35.2 ±14.8 yrs/old including 56 M(54.9%)and 46 F(45.09%).Acceptable correlation of ACD measured thru Smith method with Orbscan and Pentacam are R=0.958 and R=0.942 respectively and so Orbscan results can be used in procedures relying on ACD.
|
Conclusion:
|
ACDs measured with OrbScan is more precise than Pentacam and so can be more useful in some surgery procedures relying ACD results such as phakic IOLs and in cycloplegia contraindications.
|